ORIGINAL ARTICLE
Year : 2018  |  Volume : 7  |  Issue : 4  |  Page : 332-337

Mitigating polymerase chain reaction/amplicon contamination in a high-risk high-burden mycobacterial reference laboratory in a resource-limited setting


Intermediate Reference Laboratory, State TB Demonstration cum Training Centre, Kolkata, West Bengal, India

Date of Web Publication5-Dec-2018

Correspondence Address:
Prasanta Kumar Das
Intermediate Reference Laboratory, State TB Demonstration cum Training Centre, Kolkata - 700 010, West Bengal
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijmy.ijmy_97_18

Rights and Permissions
  Abstract 


Background: Nucleic acid amplification techniques have become important machineries in the diagnosis of several diseases in clinical laboratories. Polymerase chain reaction (PCR) contamination/amplicon contamination leading to false positivity remains a major concern in these laboratories. Prevention of these contaminations in establishing these molecular biology laboratories has been very crucial over the years. Although closed system PCRs have substantial reduction in the PCR contamination rates, the conventional probe-based hybridization methods continue to show occurrence of contamination for various reasons. The study involved checking the crucial parameters as well as the probable candidates of causing the contamination at a high-burden setting. Bringing out the most effective interventions in controlling the PCR contaminations for future endeavors stood as priority. The study explored the efficacies of different sets of interventions contributed to the process of reducing the contaminants. Methods: The detection of the contaminating PCR products or amplicons or contaminating organism is done by the genotype MTBDR plus V2 kits (Hains Life Sciences) based on DNA strip technology. Results: The pre- and post-cleaning as well as cleaning of the working surfaces was able to bring down the mean contamination percentage by 36.5%. The combined effect of cleaning the work surfaces, the automated pipetting devices, and the AC machines was able to bring down the mean contamination percentage to 53.5% reducing the contamination rate nearly to 94.6%. Conclusion: Regularly cleaning the work surfaces, the automated pipetting devices, the PCR machine, and the AC machines along with it filters and exposure of UV rays significantly lowers down the mean contamination percentage.

Keywords: Amplicon contamination, Mycobacterium tuberculosis complex, polymerase chain reaction contamination


How to cite this article:
Das PK, Ganguly SB, Mandal B. Mitigating polymerase chain reaction/amplicon contamination in a high-risk high-burden mycobacterial reference laboratory in a resource-limited setting. Int J Mycobacteriol 2018;7:332-7

How to cite this URL:
Das PK, Ganguly SB, Mandal B. Mitigating polymerase chain reaction/amplicon contamination in a high-risk high-burden mycobacterial reference laboratory in a resource-limited setting. Int J Mycobacteriol [serial online] 2018 [cited 2018 Dec 11];7:332-7. Available from: http://www.ijmyco.org/text.asp?2018/7/4/332/246923




  Introduction Top


Amplicon contamination when encountered is the most dreaded experience for any molecular biology laboratory. The public health tuberculosis (TB) laboratories at the intermediate level has probe-based molecular detection test that requires strategies to avoid polymerase chain reaction (PCR) contamination. Amplified PCR products as contaminants giving false positives may affect the patient care severely.[1],[2] Validity of an entire batch based on its detection with oligonucleotide probes is at question when the negative control shows the presence of contaminating amplicons incurring waste of resources.

The Intermediate Reference Laboratory (IRL) at Kolkata under State TB Demonstration and Training Centre (Government of West Bengal, India) has a molecular biology setup that runs line probe assay for antitubercular drugs toward the first-line (isoniazid and rifampicin) and second-line (fluoroquinolones and second-line injectables) drugs. The laboratory is well managed with internal quality check (QC) and external QC done by the National Reference Laboratories on behalf of Central TB Division, Government of India.

The laboratory encountered an occurrence of amplicon contamination that recurred in subsequent runs (January 2016). The laboratory involves a process of lysis buffer (LB)-based DNA extraction followed by PCR and hybridizing with biotinylated oligonucleotide probes on nitrocellulose strips. When a PCR is in progress, more than 10 million copies of a template DNA are produced.[3] The aerosolized amplified product may bind to the oligonucleotide probe resulting in a false positivity.

These loose amplified nucleic acid moieties are termed as amplicons which may eventually contaminate the environment. IRL Kolkata is a high-risk high-burden laboratory that involves 80–100 sample processing load daily on an average. About 24 DNA extracts reach the molecular biology setup after initial processing. This means that a billion of copies of amplified PCR products are generated on a regular basis. The huge quantum of this amplified PCR products are hence to be decontaminated regularly to avoid false positives. The PCR machine, workstations, micropipettes, and automated pipetting devices are cleaned and calibrated for ensuring valid results. The task force involved was trained by trainers from National Reference Laboratories and had a clear understanding of the amplicon control mechanisms.

The handling with proper maneuvers and commendable dexterity was ascertained. Despite these measures, contamination occurred as it was evident from the control runs. The negative control showed the presence of bands. Subsequent exercise involved conduction of onsite evaluation and addressing the shortfalls based on the experiences and observations of various studies.[4],[5],[6],[7],[8] The contamination persisted even after addressing the baseline shortfalls. Each episode of sentinel testing preceded after a thorough cleaning procedure based on the molecular detection of drug-resistant TB by line probe assay.[9] While mitigating amplicon contamination at IRL Kolkata, the following objectives were kept in consideration.

  • To establish that significant contamination has occurred
  • To find out the source of amplicon contamination
  • To find out the most contaminated room
  • To find out the area of the facility maximally affected by the amplicon contamination and its possible causes
  • The most significant interventions in controlling the contamination
  • To provide recommendations based on the inference obtained.



  Methods Top


Basic compliances and internal QCs for contamination control are in place. The entire facility design has three distinct separate units: the master mix section, the amplification or PCR unit, and the detection unit. The master mix unit involves preparation of reagents with salts, primers, and Taq polymerase. The amplification unit ensures the actual PCR. The detection unit involves hybridization with the oligonucleotide probes to detect the amplified genetic sequences. The MTBDR V2 kits of Hains Life Sciences for the detection of Mycobacterium tuberculosis complex PCR products were used.

As the laboratory was operating only one the first line of drugs, the MTBDR V2 kit for the detection of first line of drugs, this kit was taken to detect for the amplicons that are produced during PCR of the gene segments primarily responsible for resistance to the first line of drugs.

The different areas to be checked for the amplicon contaminations were placed with cryovials containing DNase, RNase-free sterile molecular water. This will help in capturing the airborne amplicons. As a comparable tool test on a paired data and there after checking on the minimal standard deviations from the mean in order to establish significant contamination at a particular site, another set of DNase RNase-free LB (Lysis Buffer) were taken and placed in the same area.

The molecular grade water (MGW) as well as the LB (sterile DNase RNase free) was then put for DNA extraction, and subsequently, amplication was done. Contaminating known set of amplicons whose primers are already present in the master mix would get amplified indicating contaminants in that particular site. Both the MGW and the LB were left open to catch up the airborne amplicons analogous to that of the settle plate technique save for the part that this is purely a molecular detection.

Testing for contaminants involved DNA extraction by LB and subjecting them for PCR. Detection of the segments was done using kit-based oligonucleotide probes on nitrocellulose paper strips. The PCR contaminants within the following gene segments were tested for [Figure 1].
Figure 1: Tested for the amplicon

Click here to view


Study design

Mechanical barriers have proved to be very effective in controlling carry-over contamination. Separate zones dividing separate areas of operation depending on the status of their cleanness provide significant control over the spread of the amplicons.[10]

Based on the above-stated conditional requisites, the molecular biology setup at the IRL Kolkata includes three separate clean rooms for designated activities: the master mix room (MMX) where the PCR reagents are prepared with the desired primers nucleotides buffer and Taq polymerase; the amplification section where PCR was performed after addition of the extracted DNA (AMP); the detection room where the PCR products are exposed to the nitrocellulose strips containing oligonucleotide probes and finally detected by a color development over the strips (DET).

Twenty-three sites are selected in each of the three sections, namely master mix section, amplification section, and detection section. A total of three runs in two batches were performed in these sites. The first batch comprised MGW and the second batch comprised LB. The cryovials containing MGW or LB were allowed to stand at those sites overnight to capture the amplicons. Significant contamination is ascertained by calculating odds ratio between the chances of occurring a contamination over not occurring a contamination in all the rooms. The population (confidence interval) was calculated taking the 414 runs (sample size 414); the frequency was taken as the numbers of occurrence of contamination, i.e., 133. The P value at 95% confidence limit was between 0.27 and 0.36.

As the source of contamination being airborne, the figuring out of the exact source becomes difficult. Sequentially, ruling out each and every step, meticulously performing a specific intervention and checking the status of contamination one after the other might provide some evidence-based insights, but being conclusive would require a significant decrease in contamination by a particular intervention at a particular site in subsequent tests. Contamination site that showed recurrent contaminations even after repeating the tests after all the intervention may suggest a site in proximity to qualifying for a source suspect.

Each section is tested for a mean percentage contamination. This is calculated out for the three runs for all the 23 sites. A mean percentage for each room is detected. The room or section showing the maximum mean percentage contamination is taken as the most contaminated room or section.

It is always not likely that the room or section with the maximum mean percentage contamination would harbor the source of contamination, but a substantial chance is an obvious phenomenon. The site scores higher percentage contamination from the mean percentage contamination with maximum variance on the higher side considering all the three runs were taken from the site that is maximally contaminated. Onsite evaluation and introducing suitable interventions would further help in finding out the cause.

The study explores 12 interventions and their effects in minimizing the mean percentage contaminations of each section.

  • Facility design
  • Workflow
  • Personal Protective Equipment adherence and controlled skillful dexterity
  • Proper disposal care and concerns
  • Proper preparation of cleaning reagents and disinfectants
  • Cleaning of floors
  • Cleaning of walls
  • Cleaning of work surfaces (before and after operations)
  • Cleaning of pipettes (before and after operations)
  • Cleaning of automated pipetting device/PCR machine
  • Use of ultraviolet (UV) radiations (intensity and effective exposure)
  • Cleaning of air-conditioners.


The effect on mean percentage contamination in all the three rooms by each of this individual is tested. [Table 1] shows that: Intervention and their contribution in controlling the mean percentage contamination. the corresponding decrease in the mean percentage contamination is seen.
Table 1: Intervention and their contribution in controlling the mean percentage contamination

Click here to view



  Results Top


A sample size of 414; 23 (sites) × 3 (runs) × 2 (fi rst with MGW and then with LB) × 3 (for three rooms) a frequency is that occurrence of contamination in 414 runs was 133 (absolute fi gures out of 414). A population confidence interval of 95% with 414 as a sample size and 133 as a frequency showed a confidence interval proportion ranging from 0.27 to 0.36. The mean contamination percentage was found to be highest in the DET. However, it was observed effective contamination control in the MMX and AMP would minimize the chances of contamination in the DET [Table 2].
Table 2: Odds ratio of the three rooms

Click here to view


As seen from the odds ratio, the chances of occurring an event of contamination are higher than an event of noncontamination, but the chance in case of the DET is high as compared to the MMX and AMP. The overall mean percentage contamination of the MMX, AMP, and DET was found to be 31%, 34%, and 39.1%, respectively. The DET shows maximum contamination of mean percentage contamination of 39.1% [Table 3].
Table 3: The mean percentage contamination in the three rooms

Click here to view


The effects of the intervention in decreasing the mean contamination percentage of the rooms showed significant reduction in contamination rates by cleaning activities. Mechanical barriers have proved to be very effective in controlling carry-over contamination. Separate zones dividing separate areas of operation depending on the status of their cleanness provides significant control over the spread of the amplicons. Unidirectional flow is maintained from the reagent preparation area to the sample preparation area, the amplification area, and finally the detection area. The chemical decontamination has been observed to be the most effective of all mechanisms. All the equipment such as pipetting devices, thermocycler, genotype blotting machines, and mini spin were cleaned with 1% hypochlorite solution (NaOCl) before and after operation. As stringent and aggressive intervention to overcome the PCR contaminant load, 4% NaOCl was used.[10]

Bleach helps in causing oxidative damage to nucleic acid.[11] Each episode of cleaning with hypochlorite was followed by cleaning the surfaces with 70% ethanol to nullify the corroding effects of bleach.[10] The pre- and post-cleaning as well as cleaning of the working surfaces was able to bring down the mean contamination percentage by 36.5%. The combined effect of cleaning the work surfaces, the automated pipetting devices, and the AC machines was able to bring down the mean contamination percentage to 53.5% reducing the contamination rate nearly to 94.6% [Table 4].
Table 4: Contribution in reducing the contamination by the sets of intervention (annexed in Table 1)

Click here to view



  Discussion Top


The IRL at Kolkata is under the State TB Demonstration and Training Center, Government of West Bengal. This is an apex institution conducting tests for presumptive multidrug-resistant and extremely drug-resistant TB cases catering all over the state West Bengal, India. The laboratory has a molecular biology setup that runs line probe assay. During January 2016, the laboratory encountered amplicons contamination in subsequent runs that led to a systemic step-by-step analysis of the source and most effective intervention in minimizing the contamination. Through a series of sentinel testing after a specific set of interventions, the contaminants started decreasing till a stage was reached where three consecutive runs showed no contamination at all.

The exercise helped in inferring some major recommendation and a singular observation that may help the laboratories performing open PCR systems. The facility is made with puff panels and epoxy flooring for ensuring smooth surfaces facilitating the cleaning process avoiding creation of any niche for accumulation of particulates. The flow of work is ensured in a sequential manner from MMX to detection room to avoid carry-over contaminants from one section to other. Reagents are prepared in a clean separate MMX before the actual polymerization at the PCR unit. This provided additional contamination control measures. Rigorous cleaning of the units before and after operation was ensured with 1% NaOCl solution followed by 70% alcohol.

Hypochlorite helped in denaturing the spilled PCR products. Seventy percent alcohol helped in the protein denaturation process toward any other contaminating organisms. Twenty-three sites from each of the rooms were selected and the cryovials containing MGW or LB were allowed to stand at those sites overnight to capture the amplicons. The odds ratio calculated out between the chances of occurring and not occurring contamination showed three major findings. It was established that significant contamination has occurred in all the three rooms.

In all the three rooms, the occurrence of contamination when tested with MGW was higher than the chances of contamination with LB. This is because each runs for LB was tested after a set of specific interventions. The higher chances of contamination occurring with the MGW showed the effectiveness of the interventions in minimizing the contamination rates. The decrease in the mean contamination percentage in these rooms after the execution of these interventions also corroborated to the fact that the interventions played effective roles in combating the contaminants. It was observed that the cleaning of the work surfaces, the automated pipetting devices, the PCR machine, and the AC machines along with it filters with 1% hypochlorite followed by 70% alcohol and exposure of UV rays significantly lowers down the mean contamination percentage. The UV irradiation sterilizes the PCR contaminants and prevents from amplification.[12],[13] Daily cleaning of the UV lights prior to and after operations helped in exposure of the desired intensity. The UV was exposed to the work surface for 20 min. Wiping the surfaces with 1–10% hypochlorite has been a very effective tool.[14] The cleaning of the working surfaces was able to bring down the contamination rate by 60%. The combined effect of the cleaning of the work surfaces, the automated pipetting devices, and the AC machines along with it filters were able to bring down the contamination rates between 90% and 95%.

As per the odds ratio calculation, the chances of contamination are 2.91 times more as compared to noncontamination in the MMX. The chances of contamination are 2.71 times more as compared to noncontamination in the AMP. The chance is maximum in the DET; this has led us to infer in our study that if contamination control is effectively done in the MMX and AMP, contamination to occur in the DET can be avoided as the contaminants of the erstwhile rooms cumulatively add up in the DET due to the unidirectional work flow.


  Conclusion and Recommendations Top


In recent times, several relevant works have been done based on line probe assay which involves molecular detection by PCR. 3.90% of mono rifampicin resistance has been shown in a study based on molecular detection.[15] The wide variation among the strains of M. tuberculosis isolated from TB cases was reported.[16] Treatment-associated death from erstwhile studies implicated the need of molecular-based rapid diagnosis.[17] Attempts have been made to probe into the genetic characterization of the gene segments used in line probe assay to mark a specific outcome based on geographic or demographic differences.[18]

All the aforesaid studies were based on PCR-based studies; thus, the relevance of controlling PCR contamination is of utmost necessity. The reference laboratories would benefit by implementing strict and specific operating and cleaning procedures.

The study showed us the effectiveness of cleaning in controlling PCR contamination in high-burden high-risk laboratory. Regularly cleaning the work surfaces, the automated pipetting devices, the PCR machine, and the AC machines along with it filters with 1% hypochlorite followed by 70% alcohol and exposure of UV rays significantly lowers down the mean contamination percentage. Room-wise installation of UV lights apart from the PCR hood, providing at least 125 μW/cm within 1 m of the thermocycler, is recommended, but care must be taken that the DNA extracts and PCR products are not exposed to it as UV irradiation may denature the enzyme Taq polymerase and oligonucleotide.[19] The intensity of the UV lights must be checked after extensive hours of its use. For G30T8 UV lamp, it has been seen that the UV output becomes 80% of that a new lamp after 8000 h of use.[20]

As there are a lot of low molecular weight moieties that float in the vicinity instead of settling down on floors or on the work surfaces, UV radiation can effectively control these contaminants. An exposure of 20 min with a desired intensity of before the operation will be a very effective tool in avoiding PCR contamination. It was found that figuring out the exact source of this airborne PCR contamination is very difficult. The contamination might have occurred during reagent preparation or addition of samples, at any step during sample maneuvers after PCR. The DET shows the highest contamination rate of mean percentage contamination of 39.1% as the sample and reagents flow from MMX to AMP followed by DET.

All the contaminants acquired in MMX and AMP reaches DET to show a maximum contamination. The odds ratio calculation also led to a singular observation reiterating the fact that cleaning of the MMX and AMP plays a pivotal role in the control of the contamination in the DET; if the contamination control is effectively done in the MMX and AMP, the chances of contamination to occur in the detection room are significantly reduced. The study showed us that absolute contamination control is hardly a reality after repeated sentinel testing that preceded after rigorous cleaning activities; there remained 5–10% chances of contamination to occur as this was observed earlier.

“It should be emphasized that despite these improvements, PCR and other amplification techniques remain susceptible to carry-over contamination. False-positive findings have been reported with all the commercially-available automated systems.[10]

Thus, it is essential to ensure clean runs in at least two negative controls (one during the preparation of reagents the other during extraction) to validate a batch of probe runs.

Acknowledgment

We are thankful to Dr. Dheeraj Tumu WHO Consultant, Revised National Tuberculosis Control Programme, West Bengal, India, for providing statistical inputs and all patients suffering and staffs contributing for their dedication.

Financial support and sponsorship

The study was supported by the Revised National Tuberculosis Control Programme, Government of India.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Molloy PJ, Persing DH, Berardi VP. False-positive results of PCR testing for Lyme disease. Clin Infect Dis 2001;33:412-3.  Back to cited text no. 1
    
2.
Patel R, Grogg KL, Edwards WD, Wright AJ, Schwenk NM. Death from inappropriate therapy for Lyme disease. Clin Infect Dis 2000;31:1107-9.  Back to cited text no. 2
    
3.
Brunstein J. Strategies for Avoiding Amplicon Contamination in the Molecular Biology Laboratory. The Primer; 2018. Available from: https://www.mlo-online.com/strategies-avoiding-amplicon-contamination-molecular-laboratory. [Last accessed on 2018 Oct 10].  Back to cited text no. 3
    
4.
Hartley JL, Rashtchian A. Dealing with contamination: Enzymatic control of carryover contamination in PCR. PCR Methods Appl 1993;3:S10-4.  Back to cited text no. 4
    
5.
Prince AM, Andrus L. PCR: How to kill unwanted DNA. Biotechniques 1992;12:358-60.  Back to cited text no. 5
    
6.
Ou CY, Moore JL, Schochetman G. Use of UV irradiation to reduce false positivity in polymerase chain reaction. Biotechniques 1991;10:442, 444, 446.  Back to cited text no. 6
    
7.
Van Pelt-Verkuil E, Belkum AV, Hays JP. Principles and Technical Aspects of PCR Amplification. 1st ed. Netherlands: Springer; 2008.  Back to cited text no. 7
    
8.
Viljoen GJ, Nel LH, Crowther JR. Molecular Diagnostic PCR Handbook. 1st ed. Netherlands: Springer; 2005.  Back to cited text no. 8
    
9.
Barnard M, Cirillo D, Feldman K, Gutierrez C, Miotto P, Parsons L, Somoskovi A. Molecular Detection of Drug Resistant Tuberculosis by Line Probe Assay. Laboratory Manual for Resource Limited Settings. Geneva, Switzerland: FIND; 2012. p. 274.  Back to cited text no. 9
    
10.
Aslanzadeh J. Preventing PCR amplification carryover contamination in a clinical laboratory. Ann Clin Lab Sci 2004;34:389-96.  Back to cited text no. 10
    
11.
Hayatsu H, Pan S, Ukita T. Reaction of sodium hypochlorite with nucleic acids and their constituents. Chem Pharm Bull (Tokyo) 1971;19:2189-92.  Back to cited text no. 11
    
12.
Dwyer DE, Saksena N. Failure of ultra-violet irradiation and autoclaving to eliminate PCR contamination. Mol Cell Probes 1992;6:87-8.  Back to cited text no. 12
    
13.
Fox JC, Ait-Khaled M, Webster A, Emery VC. Eliminating PCR contamination: Is UV irradiation the answer? J Virol Methods 1991;33:375-82.  Back to cited text no. 13
    
14.
Persing DH. Amplification product inactivation. In: Diagnostic Molecular Microbiology Principles and Applications. Washington, DC: ASM Press; 1993. p. 105-21.  Back to cited text no. 14
    
15.
Pao CC, Hor JJ, Tsai PL, Horng MY. Inhibition of in vitro enzymatic amplification reaction by ultra-violet light irradiation. Mol Cell Probes 1993;7:217-9.  Back to cited text no. 15
    
16.
Harrington BJ, Valigosky M. Monitoring ultra violet lamps in biological safety cabinets with cultures of standard bacterial strain on TSA blood agar. Lab Med 2007;38:165-8.  Back to cited text no. 16
    
17.
Sharma S, Hanif M, Chopra KK, Sharma M, Dwivedi KK, Sidiq Z, et al. Detection of multidrug resistance and extensively drug resistance among smear-negative extrapulmonary cases in a reference laboratory. Biomed Biotechnol Res J 2018;2:132-5.  Back to cited text no. 17
  [Full text]  
18.
Vashistha H, Hanif M, Chopra KK, Shrivastava D, Khanna A. Genetic polymorphism of rare mutations in Mycobacterium tuberculosis-infected patients in Delhi. Biomed Biotechnol Res J 2018;2:74-81.  Back to cited text no. 18
  [Full text]  
19.
Milanov V, Falzon D, Zamfirova M, Varleva T, Bachiyska E, Koleva A, et al. Factors associated with treatment success and death in cases with multidrug-resistant tuberculosis in Bulgaria, 2009-2010. Int J Mycobacteriol 2015;4:131-7.  Back to cited text no. 19
  [Full text]  
20.
Maurya AK, Nag VL, Kant S, Singh Kushwaha RA, Dhole TN. Genotypic characterization of rpoB, katG and inhA gene of multi drug tuberculosis isolates from extra pulmonary tuberculosis. Biomed Biotechnol Res J 2017;1:129-33.  Back to cited text no. 20
  [Full text]  


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Methods
Results
Discussion
Conclusion and R...